MEASUREMENT TRACEABILITY PROCEDURE Conformity assurance Electrical lab – Dammam Issue date: 01.04.2024 V:01 Code: TP 6.5 controlled E copy MEASUREMENT TRACEABILITY PROCEDURE TP 6.5 Author Prepared by Reviewed and approved by Name Mohamed Abdelaziz Shaik aldien Function /position Quality Lead / Senior Quality Supervisor lab Manager Signature Date All rights reserved. These materials are confidential and proprietary to Conformity Assurance. No part of these materials should be reproduced, published in any form by any means (electronic or mechanical, including photocopy or any information storage or retrieval system), nor should the materials be disclosed to third parties without the express written authorization of Conformity Assurance
MEASUREMENT TRACEABILITY PROCEDURE Conformity assurance Electrical lab – Dammam Issue date: 01.04.2024 V:01 Code: TP 6.5 controlled E copy 4. References 4.1. ISO 9000:2015 Quality Management System – Fundamentals and Vocabulary 4.2. ISO 9001:2015 Quality Management System – Requirements 4.3. ISO IEC 17025:2017 General requirements for the competence of testing and calibration laboratories 4.4. ISO/IEC Guide 99, Measurement results 5. Responsibilities 5.1. Technical manager 5.2. Lab manager 6. Process 6.1. All instruments used for tests, including equipment’s for subsidiary measurements (e.g. for environmental conditions) shall be calibrated before being put into service. 6.2. The program for calibration of instruments is designated and operated so as to ensure that calibration and measurements made by the lab are traceable to the international system of units (SI), refer to equipment procedure QP 6.4 6.3. Calibration is conducted for the equipment using SI units that provide the lab with certificates of calibration linking the calibrations to measurements made by an agency which meet CONFORMITY ASSURANCE LABORATROY requirements that had been documented under purchasing services and supplies procedure, QP 6.4 6.4. Thermometers and balances are calibrated every year unless specified other than that. 6.5. The date of calibration and the date of the next calibration that is due is included in the calibration certificates. 6.6. Calibration is verified during the performance of testing. The frequency of calibration is identified in the test method or per lab instrument working instructions. Servicing is requested from the calibration lab when data acceptance criteria for verification are not met and the analyst is unable to effect corrective action 6.7. Lab participates in inter - calibration and proficiency testing program to provide evidence of the correlation of results. 7. Flow chart NA 8. Forms /template 8.1. Metrological traceability verification report TF 6.5 -01
MEASUREMENT TRACEABILITY PROCEDURE Conformity assurance Electrical lab – Dammam Issue date: 01.04.2024 V:01 Code: TP 6.5 controlled E copy 9. Revision history The following list provides a key-word-based overview of the changes made to this QM document over time. Annex (I) : SI units Base Quantity Base unit Name Symbol name Symbol Length L, h, r meter m Mass M kilogram Km Time T Second s Electric current I. i Ampere a Temperature T Celsius c pressure P bar bar Luminous intensity lux Lux lux SN Issue No. Revision No. Date Change Training 1. 1.0 0.0 01-04-2024 Initial issue Done